Stryker Osteonics: Prosthetic Knee Joint

11 August 2021, Version 1

Abstract

This report studies the biomechanics, in particular thin-film lubrication, of an artificial knee joint. We study whether using a porous (but still stress resilient) material, or simply adding to the current bearing design a large number of small drilled holes, makes it possible to entrap a lubricating fluid, in a way similar to what the living tissue does. The report concludes that such a design is indeed possible and may lead to less friction in low-load situation; but that the high-load situation is less clear. Also, the report considers only sliding movement of the bearing, not a combination of sliding and rolling.

Content

Comments

Comments are not moderated before they are posted, but they can be removed by the site moderators if they are found to be in contravention of our Commenting Policy [opens in a new tab] - please read this policy before you post. Comments should be used for scholarly discussion of the content in question. You can find more information about how to use the commenting feature here [opens in a new tab] .